X-RAY Generator: HIGH FREQUENCY MONOBLOC
PANEL-PC WORkstation
Maximum power: 32 kW (320 mA - 100 kV @ 100 ms)
Maximum voltage: 125 kV
Maximum mAs: 320 mAs
Maximum mA: 400 mA
Monobloc thermal capacity: 600 kJ (800 KHU)
Monobloc continuous thermal dissipation: 55 W
Frequency: 40 kHz
kV ripple: < 2%
Leakage radiation (IEC 601-1-3): < 0,5 mGy/h
Total filtration: > 2,5 mm Al @ 75 kV
CPU Processor: Intel® CoreTM i5-4402E dual-core 1.6 GHz
Memory: 4GB DDR3L SO-DIMM
Hard Disk Drive: 110 GB
Peripherals: DVD-RW
Operating System: Windows® 7 Professional
Local storage of uncompressed images: 18.000
Input / Output external interface: 1 x USB 2.0
1x ethernet RJ45
Backup battery capacity: 7800 mAh Li-Ion
X-RAY TUBE
50kHz
PANEL PC-Display
Resolution: 1280 x 1024 (SXGA) with 16.7M colors
Touch screen: Analogue resistive touch screen, 80% Light transparency (typical)
Luminance: 350 cd/m2 (typical)
Contrast ratio: 1000:1 (typical)
Backlight type: L.E.D
View angle (U/D/R/L): 80° / 80° / 85° / 85° (typical)
Focal spot size: 0,6 mm - 1,3 mm
Maximum voltage: 130 kV
Anode speed: 3000 r.p.m.
Anode material: RTM
Anode angle: 15°
Anode heat storage capacity: 80 kJ (107 kHU)
Maximum continuous anode heat dissipation: 320 W
IMAGING SYSTEM
Generator controls: Integrated into graphical interface.
4 patient sizes selection.
Up to 40.000 pre-programmable settings for anatomical part
and kind of projection.
Working technique: 3 points by setting kV, mA and ms
2 points by setting kV and mAs
Exposure time: 3 ms ÷ 6300 ms (34 steps)
kV range: 40 kV ÷ 125 kV (1 kV step)
mA range: 25 mA ÷ 400 mA (13 steps)
mAs range: 0,1 mAs ÷ 320 mAs (36 steps)
Exposure control: Handswitch with 3.8 m extendable cable
FixRad Software: Software to capture, view and process radiological images.
Elaboration tools:
-
Rotation
-
Inversion
-
Harmonization
-
Spatial Filters
-
Zoom
-
Graphic functions (Text, Angle, Calibration, Grid, Ruler, Black mask, Image statistics, Frame, Arrow)
-
LUT management
-
Electronic collimators selection
-
Detector magnification
-
Virtual keyboard
-
Automatic Recognition of the collimator
-
Grid suppression
Dicom print: Image printing interface
Dicom Worklist: Managing the daily list of patients
FLAT PANEL DETECTOR
Sensor protection plate: Carbon fiber plate
Conversion screen: Cesium Iodide (Csl) with amorphous Silicon (a-Si) photodiode
Active area: 350 (H) x 426 (V) mm
Pixel matrix: 2448 (H) x 2984 (V) pixel
Pixel pitch: 143 μm
Cycle time: 6 s shot to shot
Limiting resolution: 3.7 Lp/mm (typical)
MTF (2.0 Lp/mm, 70 kVp, 1×1): 36 % (typical)
DQE (0), Quantum - Limited: > 70 % (typical)
A/D conversion: 16 bit
Energy range: 40 ÷ 150 kVp
Maximum entrance dose: 4 mR/frame (low - gain linear output range)
Image acquisition exposure period: 500 ms standard (variable between 500 ms to 4000 ms)
Overall dimensions: 384 x 460 x 15 mm
ENVIRONMENTAL
-
Under delivery and stock
Temperature: - 15 to 55 °C
Humidity: 10 to 90 % (non-condensing)
Pressure: 50 to 106 kPa
-
Under operating
Temperature: + 10 to 35 °C
Humidity: 10 to 85 % (non-condensing)
OPTIONAL
DICOM Spooler: Transmits images to the remote DICOM device via the network
DICOM CD burning: Copy the images of DICOM Archive onto CD-ROM
DICOM PRINT and Film editor: Composition of the film print
DICOM Query/Retrieve: Server Query
DICOM Storage Commitment: Final storage of the images on another device
DICOM MPPS: Information about the images that are transferred to the server
DICOM DOSE SR: Structured Dose Report
OPTIONAL
-
Flat Panel antishock frame with handle for Toshiba Flat Panel Detector
-
Flat Panel antishock frame with handle, for Toshiba Flat Panel Detector complete with grid 215l/inch R=8:1
SAFETY AND PROTECTION
Over current protection
Overvoltage protection
X-ray tube overload protection
Monobloc kHU automatic survey
Errors description
COLLIMATOR
Adjustment from 0x0 to 43x43 cm at 1 m focal distance (FFD)
High intensity lamp (160 lux) for X-ray field simulation and automatic switch Off programmable by
software from 5 s to 120 s.
Retractable tape measure for FFD measurement
± 90° collimator rotation
MOBILE STAND
Antistatic rubber wheels
Dead-man operated parking brake
Variable focus-floor distance from 450 mm to 2000 mm
Monobloc yoke support rotation: +90° / -90°
Monobloc rotation in the yoke: -35° / +190°
Overall dimensions in transport position 109 × 68 × 158 (H) cm
Weight (without Flat Panel Detector): ≈ 215 kg
POWER SUPPLY MANAGEMENT
Voltage: 230 Vac single phase
Frequency: 50 Hz / 60 Hz
Nominal current: 5 A
Momentary current: 32 A
Total line resistance: 0,4 Ω
Automatic line compensation: ± 10%
On request:
Voltage: 115 Vac single phase
Frequency: 50 Hz / 60 Hz
Nominal current: 9,5 A
Momentary current: 46 A
Automatic line compensation: ± 10%
PRODUCT FEATURES
30 kW output power, from any standard 3.5 kW Power Supply socket
35 x 43 cm portable CsI Flat Panel Detector
The 63kW generator makes a difference, the generator assures the best image in every operating condition along with reduced patient close and extended x-ray tube life-span.
Computer System processor: Intel Core Quad Processor i5-6500 (3.20 GHz)
Memory: 4GB DDR3L
Hard Disk: 1Tb SATA (7200rpm)
Optical Drive: DVD-RW disc burner
19” Touch Screen Monitor for radiographic controls and image viewing
The design is a fixed digital x-ray machine, but it can be fitted into limited spaces such as buses and trucks.
Full DICOM connectivity and CD/DVD recording
Our generator provide the right amount of penetration, reducing exposure times and rtetakes. Automatic exposure control (AEC) enables auto-selection of radiographic factors, saves time, eliminates retakes, increases diagnostic capability, and lowers the radiation dose (optional).
Intuitive software for high resolution acquisition and post-processing
Flat panel detector improves the workflow, exam speed, and provides comfort with efficiency. A-si detector with CSI screen, excellent spatial resolution. Active area: 17x17"